학술논문

A Low-Cost Handheld Reconfigurable Impedimetric Readout System for Diagnostics of Viral Infections
Document Type
Periodical
Source
IEEE Transactions on Instrumentation and Measurement IEEE Trans. Instrum. Meas. Instrumentation and Measurement, IEEE Transactions on. 72:1-8 2023
Subject
Power, Energy and Industry Applications
Components, Circuits, Devices and Systems
Sensors
Electrodes
Impedance
COVID-19
Integrated circuits
Antibodies
Impedance measurement
Biosensor
electronic readout system
impedance measurement
SARS-CoV-2
Language
ISSN
0018-9456
1557-9662
Abstract
This article presents a portable impedimetric biosensor for detecting infectious diseases, such as SARS-CoV-2 infections. A bio-ready sensing electrode functionalized with SARS-CoV-2 nucleocapsid antibody was employed to quantitatively convert the concentration of nucleocapsid protein (N-protein) into impedance changes. In this article, we proposed a readout system with a dynamic input range of 200- $\Omega $ –1- $\text{M}\Omega $ magnitude and 0°–180° phase. The resolution of this device is 1% and 6.5° for measuring the magnitude and phase, respectively. Herein, we demonstrate and discuss the proposed system’s functionality, sensitivity, and selectivity using the clinical swab samples. As per these results, this readout system is suitable for the detection of N-protein ranging up to 10000 pg/mL with a resolution of 56 fg/mL. The proposed impedimetric sensing system can be adopted for the detection of infectious diseases in the future. This low-cost (< $\$ $ 80) device using off-the-shelf is a unique candidate for batch production purposes during urgent pandemic situations.