Noninvasive Cardiac Output and Central Systolic Pressure From Cuff-Pressure and Pulse Wave Velocity
Document Type
Periodical
Source
IEEE Journal of Biomedical and Health Informatics IEEE J. Biomed. Health Inform. Biomedical and Health Informatics, IEEE Journal of. 24(7):1968-1981 Jul, 2020
Subject
Bioengineering
Communication, Networking and Broadcast Technologies
Components, Circuits, Devices and Systems
Computing and Processing
Signal Processing and Analysis
Sensitivity
Brachytherapy
Mathematical model
Hemodynamics
Pressure measurement
Arteries
Noninvasive
cardiac output
central pressure
1-D model
patient-specific models
optimization methods
Language
ISSN
2168-2194
2168-2208
Abstract
Goal: We introduce a novel approach to estimate cardiac output (CO) and central systolic blood pressure (cSBP) from noninvasive measurements of peripheral cuff-pressure and carotid-to-femoral pulse wave velocity (cf-PWV). Methods: The adjustment of a previously validated one-dimensional arterial tree model is achieved via an optimization process. In the optimization loop, compliance and resistance of the generic arterial tree model as well as aortic flow are adjusted so that simulated brachial systolic and diastolic pressures and cf-PWV converge towards the measured brachial systolic and diastolic pressures and cf-PWV. The process is repeated until full convergence in terms of both brachial pressures and cf-PWV is reached. To assess the accuracy of the proposed framework, we implemented the algorithm on in vivo anonymized data from 20 subjects and compared the method-derived estimates of CO and cSBP to patient-specific measurements obtained with Mobil-O-Graph apparatus (central pressure) and two-dimensional transthoracic echocardiography (aortic blood flow). Results: Both CO and cSBP estimates were found to be in good agreement with the reference values achieving an RMSE of 0.36 L/min and 2.46 mmHg, respectively. Low biases were reported, namely −0.04 ± 0.36 L/min for CO predictions and −0.27 ± 2.51 mmHg for cSBP predictions . Significance: Our one-dimensional model can be successfully “tuned” to partially patient-specific standards by using noninvasive, easily obtained peripheral measurement data. The in vivo evaluation demonstrated that this method can potentially be used to obtain central aortic hemodynamic parameters in a noninvasive and accurate way.