A large area MicroGap Chamber with two-dimensional read-out
Document Type
Periodical
Source
IEEE Transactions on Nuclear Science IEEE Trans. Nucl. Sci. Nuclear Science, IEEE Transactions on. 43(3):1237-1242 Jun, 1996
Subject
Nuclear Engineering
Bioengineering
Strips
Current measurement
Charge measurement
Particle beam measurements
Collision mitigation
Anodes
Sampling methods
Cathodes
Detectors
Gain measurement
Language
ISSN
0018-9499
1558-1578
Abstract
A 10/spl times/10 cm/sup 2/ MicroGap Chamber with a small stereo angle read-out has been developed for the inner tracking system of CMS. The anode strips are 9 /spl mu/m wide with a pitch of 200 /spl mu/m. The sampling pitch of the back cathode is 400 /spl mu/m at a stereo angle of 50 mrad. Both coordinates are read-out from the same side of the detector substrate. Measurements of gain stability, fraction of induced charge on the back electrode, uniformity of response along the strip and rate capability have been performed in laboratory, using X-ray sources. Detection efficiency, spatial resolution, charge and space correlation have been measured with minimum ionizing particle beams at CERN.