Performance Evaluation of Liquinert-Processed CeBr₃ Crystals Coupled With a Multipixel Photon Counter
Document Type
Periodical
Source
IEEE Transactions on Nuclear Science IEEE Trans. Nucl. Sci. Nuclear Science, IEEE Transactions on. 67(6):988-993 Jun, 2020
Subject
Nuclear Engineering
Bioengineering
Hafnium
DH-HEMTs
Energy resolution
Cerium (III) bromide (CeBr₃)
coincidence resolving time (CRT)
energy resolution
multipixel photon counter (MPPC)
Language
ISSN
0018-9499
1558-1578
Abstract
Ce-doped scintillators are capable of shortening the time constant as the luminescence is caused by 5d $\to$ 4f transition of Ce 3+ . A scintillator with a short-time constant is suitable for time-of-flight (TOF)-positron emission tomography (PET). In this article, we conducted an evaluation of the energy resolution and coincidence resolving time (CRT) of Liquinert-processed cerium (III) bromide (CeBr 3 ), which possesses a short-time constant and has a high light yield, using multipixel photon counter (MPPC) as a photon detector. The results indicate that an energy resolution of 5.02% (full width at half maximum (FWHM)) can be obtained from a 5-mm $\times \,\, 5$ -mm $\times \,\, 5$ -mm-CeBr 3 cube coupled with a 6.0-mm $\times \,\, 6.0$ -mm-pitch MPPC. Furthermore, a CRT of 97.0 ps (FWHM) was achieved using a 3-mm $\times \,\, 3$ -mm $\times \,\, 1$ -mm-CeBr 3 array coupled with a 3.0-mm $\times \,\, 3.0$ -mm-pitch through-silicon-via MPPC array, which has been proven effective for TOF-PET and Compton imaging. Moreover, a CRT of 198.7 ps (FWHM) was obtained using a current comparing type time-over-threshold circuit as a front-end readout application. We also determined that the energy resolution is independent of the shaping time but depends on the temperature. In addition, the CRT was found to be dependent on the frequency bandwidth of the signal measuring instrument.