학술논문

Large Matrix Array Aperture for 3D Vascular Imaging Capture
Document Type
Conference
Source
2022 IEEE International Ultrasonics Symposium (IUS) Ultrasonics Symposium (IUS), 2022 IEEE International. :1-4 Oct, 2022
Subject
Bioengineering
Fields, Waves and Electromagnetics
Signal Processing and Analysis
Transducers
Three-dimensional displays
Ultrasonic imaging
Ultrasonic variables measurement
Wires
Imaging
Ultrasonic transducer arrays
Matrix array
transducer
reconfigurable
3D
ASIC integration
Language
ISSN
1948-5727
Abstract
Three-dimensional ultrasound has initially been used to address volumetric imaging for diagnostic purposes and represents the leading-edge technological orientation in both transducer and IC (integrated circuit) architecture and design. However, new applications are coming up like biomarker measurements, preoperative navigation, real time surgery guidance or therapeutic procedures where 3D ultrasound modalities are key but their design objectives may need to be thought outside 3D echocardiography and radiology technological trade-offs. For those new applications, system architectures would need less complexity and imaging performances enabling easier hardware reconfigurability tailored to application-oriented imaging. This paper presents an ongoing development where a large matrix transducer has been assembled with multiple ASIC dies in a reconfigurable way. The transducer has a central frequency of 8MHz, a square pitch of 150µm× 150µm capable to fully image the upper carotid window thanks to a large aperture of 80×240 elements, resulting in a transducer active footprint of 12×36 square millimeters.