학술논문

Mechanical comparison between two titanium micro membranes for in vivo pressure monitoring
Document Type
Conference
Source
2020 Symposium on Design, Test, Integration & Packaging of MEMS and MOEMS (DTIP) Design, Test, Integration & Packaging of MEMS and MOEMS (DTIP), 2020 Symposium on. :1-6 Jun, 2020
Subject
Components, Circuits, Devices and Systems
Micromechanical devices
Fabrication
In vivo
Sensitivity
Transducers
Titanium
Packaging
medical implants
MEMS
pressure sensor
bulk titanium
micromachined membrane
embossed membrane
Language
Abstract
We report the fabrication and characterization of two kinds of titanium membranes made of bulk titanium. Both membrane designs are compared mechanically to define the most suited for the targeted application. These membranes are future parts of Ti-based MEMS pressure sensors for active biomedical implants. As pressure is known to be linked to patient’s health for numerous diseases (glaucoma, heart failure, hydrocephalus...), in vivo pressure monitoring appears as a promising way to help medical doctors to define their treatments. Yet, most of current methods used to monitor pressure are still routinely invasive, which increases the risk of infection. The use of Ti as a mechanical transducer appears as a promising approach to tackle the lack of biocompatibility of standard semiconductor materials such as silicon, which requires a complex, bulky and costly packaging.