Prior Image-Constrained Iterative Reconstruction With Adaptive Step Size for Limited-Angle CBCT
Document Type
Periodical
Source
IEEE Transactions on Instrumentation and Measurement IEEE Trans. Instrum. Meas. Instrumentation and Measurement, IEEE Transactions on. 73:1-13 2024
Subject
Power, Energy and Industry Applications
Components, Circuits, Devices and Systems
Image reconstruction
Computed tomography
Radiation therapy
TV
Planning
Compressed sensing
Imaging
Adaptive step size
iterative reconstruction
limited-angle CT reconstruction
prior image constrained
Language
ISSN
0018-9456
1557-9662
Abstract
Cone-beam computed tomography (CBCT) has been widely used in image-guided radiotherapy (IGRT). In order to avoid the possible collisions of the moving gantry with patients and devices, the limited-angle scanning protocol is often adopted for CBCT. However, the missing views of projection data will induce severe wedge artifacts to the reconstructed images, having a negative influence on the following therapeutic procedures. Compared to limited-angle CBCT images, the planning CT images (pCT) acquired earlier for the same patient are artifact-free, which has the potential to improve the quality of limited-angle CBCT. In this context, this article proposes the prior image-constrained adaptive step size iterative reconstruction (PICAS) method. PICAS builds on prior image-constrained compressed sensing (PICCS) but is improved after taking into account the characteristics of IGRT. In PICAS, the high-quality pCT images are regarded as prior images to enhance the performance in artifact removal. To address the mismatch between pCT images and CBCT images, the pCT images are reprojected and reconstructed using the same imaging geometry as CBCT. Then, a transformation matrix between the pseudo-limited-angle pCT images and limited-angle images is obtained and it is further applied to the high-quality pCT images to generate the artifact-free prior images. In addition, an adaptive gradient descent (GD) optimization based on convex set projection and Lipschitz constant is adopted to accelerate the convergence of the algorithm. The proposed PICAS algorithm has been evaluated on the real data from different parts (head, lung, and abdomen) and under the limited-angle configurations with different scanning ranges (80°, 100°, and 120°). Qualitative and quantitative results demonstrate that PICAS has the potential to quickly reconstruct high-quality CBCT images in limited-angle scanning scenes, and thereby improve the accuracy of radiotherapy and shorten the time of surgery.