Optimization of pulse-height sharing for use below 150 keV in long bars of NaI(Tl)
Document Type
Conference
Source
2011 IEEE Nuclear Science Symposium Conference Record Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE. :440-443 Oct, 2011
Subject
Nuclear Engineering
Bioengineering
Communication, Networking and Broadcast Technologies
Components, Circuits, Devices and Systems
Computing and Processing
Signal Processing and Analysis
Image resolution
Optical imaging
Energy measurement
Language
ISSN
1082-3654
Abstract
In this work we present an optimization of the energy and position resolutions for use in gamma-ray interactions in a long bar of NaI(TI). This work focuses on tuning the effective optical attenuation coefficient to approximately the same length of the bar. This is achieved by altering the surface of the scintillator in a systematic manner. We present data obtained in the laboratory from measurements of eighteen bars of NaI(TI) that have been manufactured with various treatments of the surface. This process results in an improvement in position resolution by a factor of four over a polished bar with only a corresponding 20% decrease in energy resolution.