학술논문

Precision measurement technique of photon detection efficiency of Silicon photomultiplier
Document Type
Conference
Source
2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC) Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE. :1-3 Oct, 2013
Subject
Nuclear Engineering
Voltage measurement
Wavelength measurement
Detectors
Photonics
Photodiodes
NIST
Silicon
Language
ISSN
1082-3654
Abstract
Silicon photomultiplier (SiPM) is the next generation sensor that is available for single photon counting. We measured the photon detection efficiency (PDE) for the SiPM in the 400 to 800 nm range and over bias voltage up to 4V using photon counting method. The experimental setup consists of multi-wavelength LEDs, a monochromator, two 2-inch integrating spheres, a NIST calibrated reference photodiode and 1×1mm SiPM (1600 micro-pixel). We use the two integrating spheres system to measure PDE measurement. The advantage of this system that we can control light intensity to match dynamic range between the reference photodiode and the SiPM in low light condition. We also calculate precisely the irradiance on the SiPM through the ray-tracing simulation of the experimental setup. We present the results of the PDE measurement as well as the measurement technique