Performance of a lead radiator, gas tube calorimeter
Document Type
Periodical
Source
IEEE Transactions on Nuclear Science IEEE Trans. Nucl. Sci. Nuclear Science, IEEE Transactions on. 36(1):86-89 Feb, 1989
Subject
Nuclear Engineering
Bioengineering
Wires
Strips
Electrons
Energy measurement
Area measurement
Laboratories
Sampling methods
Anodes
Aluminum
Calibration
Language
ISSN
0018-9499
1558-1578
Abstract
The design and performance of a 4.2-radiation-length, lead-sandwich, gas tube hodoscope are discussed. The device, measuring 1*2 m/sup 2/ in area and 12 cm in depth, was used in Fermi National Accelerator Laboratory experiment 705. Multiple samplings of anode wires situated within three-walled aluminum tubes were used to generate an X coordinate; similarly, capacitively coupled copper-clad strips were ganged together to yield a Y coordinate. The results reviewed of an analysis of electron calibration data taken during a recent six-month running period are reviewed. The position resolution (in millimeters) is seen to be 0.8+3.3/ square root E+31/E for the 9.92 mm-spaced wires and 0.6+3.2/ square root E+32 for the 12.5-mm strips, where E represents the electron beam energy in GeV. The correlation between energy as measured by the wires and energy as measured by the strips is fairly tight, with an asymmetry RMS (root mean square) for 30-GeV electrons of about 4%.ETX