Comparative analysis of hematocrit measurements by dielectric and impedance techniques
Document Type
Periodical
Source
IEEE Transactions on Biomedical Engineering IEEE Trans. Biomed. Eng. Biomedical Engineering, IEEE Transactions on. 52(3):549-552 Mar, 2005
Subject
Bioengineering
Computing and Processing
Components, Circuits, Devices and Systems
Communication, Networking and Broadcast Technologies
Dielectric measurements
Impedance measurement
Blood
Plasma measurements
Frequency measurement
Permittivity
Conductivity measurement
Electric variables measurement
Biomedical engineering
Instruments
Beta relaxation
blood
cell membrane
erythrocytes
Language
ISSN
0018-9294
1558-2531
Abstract
In a previous paper, a new dielectric technique was used to estimate hematocrit (HTC) in extracorporeal blood circulation systems, independently of plasma conductivity or osmolarity. Although many impedance techniques have been formerly proposed in the literature, none has been evaluated against plasma conductivity and osmolarity. Herein, we estimate HTC based on permittivity changes and also with other four techniques found in the literature. Besides, the error incurred in each is also studied when plasma conductivity and osmolarity changed as much as 1 mS/cm and 50 mOsm/kg, respectively. The dielectric (permittivity) technique has an error close to 5.4%, while the others showed both tendencies, i.e., lower error (2.5%, two of them) and higher error (8.6% and 16.3%, the other two). The dielectric technique, even though did not produce the lowest error, provides a well-described physical model along with simple instrumentation.