학술논문

Flexible Needle Bending Model for Spinal Injection Procedures
Document Type
Periodical
Source
IEEE Robotics and Automation Letters IEEE Robot. Autom. Lett. Robotics and Automation Letters, IEEE. 8(3):1343-1350 Mar, 2023
Subject
Robotics and Control Systems
Computing and Processing
Components, Circuits, Devices and Systems
Needles
Robots
Biological system modeling
Stress
Biological tissues
Bending
Robot sensing systems
Medical robots and systems
surgical robotics: steerable catheters/needles
Language
ISSN
2377-3766
2377-3774
Abstract
An in situ needle manipulation technique used by physicians when performing spinal injections is modeled to study its effect on needle shape and needle tip position. A mechanics-based model is proposed and solved using finite element method. A test setup is presented to mimic the needle manipulation motion. Tissue phantoms made from plastisol as well as porcine skeletal muscle samples are used to evaluate the model accuracy against medical images. The effect of different compression models as well as model parameters on model accuracy is studied, and the effect of needle-tissue interaction on the needle remote center of motion is examined. With the correct combination of compression model and model parameters, the model simulation is able to predict needle tip position within submillimeter accuracy.