Evaluation of an MRI-Guided PET Image Reconstruction Approach With Adaptive Penalization Strength
Document Type
Periodical
Source
IEEE Transactions on Radiation and Plasma Medical Sciences IEEE Trans. Radiat. Plasma Med. Sci. Radiation and Plasma Medical Sciences, IEEE Transactions on. 8(3):277-286 Mar, 2024
Subject
Nuclear Engineering
Engineered Materials, Dielectrics and Plasmas
Bioengineering
Computing and Processing
Fields, Waves and Electromagnetics
Image reconstruction
Magnetic resonance imaging
Spatial resolution
Reconstruction algorithms
Positron emission tomography
Image quality
Biomedical imaging
Adaptive spatially variant hyperparameter
[¹⁸F]-FDG-epilepsy
maximum a posteriori (MAP)
MRI-guided (MRIg)
positron emission tomography (PET)
tau-imaging
Language
ISSN
2469-7311
2469-7303
Abstract
MRI-guided (MRIg) positron emission tomography (PET) reconstruction can potentially reduce noise and increase spatial resolution compared to standard clinical ordered-subsets expectation-maximization (OSEM) image quality. However, to adjust for the desired image quality, the balance between measured data and prior information usually requires manual tuning. This work presents an adaptive method to automatically control the influence of the magnetic resonance imaging (MRI) information on the PET emission data using maximum a posteriori (MAP) image reconstruction, robust against a wide range of counts. The method was evaluated on different static brain PET datasets using [18F]-FDG, [18F]-Florbetapir and [11C]-PiB, acquired in a simultaneous PET/MRI scanner and a PET/CT scanner, followed by an MRI scan. Noise in gray and white matter was measured for a wide range of statistics. Furthermore, noise and quantification accuracy were analyzed in different cortical and subcortical brain regions with different levels of tracer uptake, and at different levels of counts. Results demonstrated consistent improved image quality in terms of noise and spatial resolution with MRI-guided MAP PET (MRIg-MAP) reconstruction compared to OSEM. Additionally, it was shown that the number of collected counts could be reduced by ~1.6– $2.3\times $ using MRIg-MAP reconstruction compared to OSEM, without increasing the noise significantly, either by reducing the scan time or injected activity. In conclusion, we presented a novel method to adaptively balance the influence of the anatomical information on the emission data for MRIg-MAP reconstruction, which showed image quality improvements compared to OSEM for different radiotracers, at different levels of counts, and applicable to simultaneous and sequential PET-MRI scans.