학술논문

Development of a Novel Wearable Ring-Shaped Biosensor
Document Type
Conference
Source
2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) IEEE Engineering in Medicine and Biology Society (EMBC), 2018 40th Annual International Conference of the. :3750-3753 Jul, 2018
Subject
Bioengineering
Stress
Heart rate variability
Standards
Prototypes
Sensors
Gold
Fingers
Language
ISSN
1558-4615
Abstract
We report on the preliminary results obtained out of a wearable module designed to be encompassed within a ring-shaped system aimed at providing healthcare services. The module is composed of two sensors for the measuring of Galvanic Skin Response (GSR) and Heart Rate Variability (HRV). A first device validation was carried out by involving four subjects who were asked to perform tasks providing different stress-related statuses. A comparison of physiological parameters measured by the module with those measured by a commercial HRV-GSR sensor chosen as gold standard was made. Two out of the three HRV parameters and all of the GSR parameters measured with the module resulted consistent (mostly differing less than 10%) with the same parameters measured by the gold standard. The work reported in this paper set a milestone for the realization of a system exploiting sensor fusion to provide active ageing, stress detection, activity recognition and e-health services has been achieved.