학술논문

Validation of a wrist monitor for accurate estimation of RR intervals during sleep
Document Type
Conference
Source
2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE. :5493-5496 Jul, 2013
Subject
Engineered Materials, Dielectrics and Plasmas
Monitoring
Wrist
Biomedical monitoring
Electrocardiography
Sleep apnea
Time series analysis
Skin
Language
ISSN
1094-687X
1558-4615
Abstract
While the incidence of sleep disorders is continuously increasing in western societies, there is a clear demand for technologies to asses sleep-related parameters in ambulatory scenarios. The present study introduces a novel concept of accurate sensor to measure RR intervals via the analysis of photo-plethysmographic signals recorded at the wrist. In a cohort of 26 subjects undergoing full night polysomnography, the wrist device provided RR interval estimates in agreement with RR intervals as measured from standard electrocardiographic time series. The study showed an overall agreement between both approaches of 0.05 ± 18 ms. The novel wrist sensor opens the door towards a new generation of comfortable and easy-to-use sleep monitors.