Results of an International Comparison of Instrument Current Transformers up to 10 kA at 50 Hz Frequency
Document Type
Conference
Source
2018 Conference on Precision Electromagnetic Measurements (CPEM 2018) Precision Electromagnetic Measurements (CPEM 2018), 2018 Conference on. :1-2 Jul, 2018
Subject
Fields, Waves and Electromagnetics
Current transformers
Current measurement
Laboratories
Uncertainty
Standards
Atmospheric measurements
Particle measurements
Current ratio
current transformers (CT)
current measurement
instrument transformers
ratio error
phase displacement
comparison
high current
Language
ISSN
2160-0171
Abstract
Traceability of the current ratio is crucial for the measurement of electrical energy in revenue metering. A comparison of the AC current ratio was therefore performed within EURAMET in the time period 2012 - 2016, using a precision CT as the traveling device. The Czech Metrology Institute (CMI) as coordinator processed the measurement results of the 15 European participating laboratories. The comparison of the results for transformer ratios of (4, 5, 6, 8, 10) kA/5 A at 15 VA burden and (4 and 10) kA/5 A ratios at 5 VA burden indicates good agreement between the participating laboratories. The main differences are found for phase displacement, at least in part due to instability of the traveling standard.