In vivo evaluation of safety and performance of a tapered nitinol venous stent with inclined proximal end in an ovine iliac venous model
Document Type
article
Source
Scientific Reports, Vol 14, Iss 1, Pp 1-14 (2024)
Subject
Iliac vein
Iliocaval venous obstruction
Venous insufficiency
Venous stent
Medicine
Science
Language
English
ISSN
2045-2322
Abstract
Abstract A tapered stent with inclined proximal end is designed for fitting the iliac anatomically. The aim of the present study was to evaluate the safety and performance of the new stent in ovine left iliac veins. The experiment was performed in 30 adult sheep, and one nitinol-based VENA-BT® iliac venous stent (KYD stent) was implanted into each animal’s left common iliac vein. Follow-up in all sheep consisted of angiographic, macroscopic, and microscopic examinations at Day 0 ( 0.05), indicating no serious lumen loss occurred during the follow-up period. Common iliac venous pressure was further measured and the results further indicated the lumen patency at follow-up. Histological examinations indicated that no vessel injury and wall rupture, stent damage, and luminal thrombus occurred. There was moderate inflammatory cell infiltration around the stent in Day-0 and Day-30 groups with the average inflammation score of 2.278 and 2.167, respectively. The inflammatory reaction was significantly reduced in Day-90, Day-180 and Day-360 groups and the average inflammation scores were 0.9444 (p