Beam quality and the mystery behind the lower percentage depth dose in grid radiation therapy
Document Type
article
Source
Scientific Reports, Vol 14, Iss 1, Pp 1-13 (2024)
Subject
Grid therapy
Small field dosimetry
Spectrometry
Monte Carlo
Linac
Medicine
Science
Language
English
ISSN
2045-2322
Abstract
Abstract Grid therapy recently has been picking momentum due to favorable outcomes in bulky tumors. This is being termed as Spatially Fractionated Radiation Therapy (SFRT) and lattice therapy. SFRT can be performed with specially designed blocks made with brass or cerrobend with repeated holes or using multi-leaf collimators where dosimetry is uncertain. The dosimetric challenge in grid therapy is the mystery behind the lower percentage depth dose (PDD) in grid fields. The knowledge about the beam quality, indexed by TPR20/10 (Tissue Phantom Ratio), is also necessary for absolute dosimetry of grid fields. Since the grid may change the quality of the primary photons, a new $${\mathbf{k}}_{\mathbf{q},{\mathbf{q}}_{0}}$$ k q , q 0 should be evaluated for absolute dosimetry of grid fields. A Monte Carlo (MC) approach is provided to resolving the dosimetric issues. Using 6 MV beam from a linear accelerator, MC simulation was performed using MCNPX code. Additionally, a commercial grid therapy device was used to simulate the grid fields. Beam parameters were validated with MC model for output factor, depth of maximum dose, PDDs, dose profiles, and TPR20/10. The electron and photon spectra were also compared between open and grid fields. The dmax is the same for open and grid fields. The PDD with grid is lower (~ 10%) than the open field. The difference in TPR20/10 of open and grid fields is observable (~ 5%). Accordingly, TPR20/10 is still a good index for the beam quality in grid fields and consequently choose the correct $${\mathbf{k}}_{\mathbf{q},{\mathbf{q}}_{0}}$$ k q , q 0 in measurements. The output factors for grid fields are 0.2 lower compared to open fields. The lower depth dose with grid therapy is due to lower depth fluence with scatter radiation but it does not impact the dosimetry as the calibration parameters are insensitive to the effective beam energies. Thus, standard dosimetry in open beam based on international protocol could be used.