학술논문

Visual versus fully automated assessment of left ventricular ejection fraction
Document Type
article
Source
Avicenna Journal of Medicine, Vol 8, Iss 02, Pp 41-45 (2018)
Subject
auto-ejection fraction
ejection fraction
m-mode
tissue tracking
visual assessment
Medicine
Language
English
ISSN
2231-0770
2249-4464
Abstract
Introduction: The aim of this study is to compare three different echocardiographic methods commonly used in the assessment of left ventricle (LV) ejection fraction (EF). Methods: All patients underwent full echocardiography including LVEF assessed using M-mode, automated EF (Auto-EF), and visual estimation by two readers. Results: We enrolled 268 patients. Auto-EF measurement was feasible in 240 (89.5%) patients. The averaged LVEF was (52% ± 12) with the visual assessment, (51% ± 11) with Auto-EF and (57% ± 13) with M-mode. Using Bland-Altman analysis, we found that the difference between the mean visual and the Auto-EF was not significant (−0.3% [−0.5803–0.0053], P = 0.054). However, the mean EF was significantly different when comparing visual versus M-mode and Auto-EF versus M-mode with the mean differences: (−2.4365 [−2.9946–−1.8783], P < 0.0001) and (−2.1490 [−2.7348–−1.5631], P < 0.0001) respectively. Inter-observer variability analysis of the visual EF assessment between the two readers showed that intraclass correlation coefficient was 0.953, (95% confidence interval: 0.939–0.965, P < 0.0001), with excellent correlation between the two readers: R = 0.911, P < 0.0001). Conclusion: The two-dimensional echocardiographic methods using Biplane Auto-EF or visual assessment were significantly comparable, whereas M-mode results in an overestimation of the LVEF.