Comparison of accuracy for hip-knee-ankle (HKA) angle by X-ray and knee motion analysis system and the relationships between HKA and gait posture
Document Type
article
Source
BMC Musculoskeletal Disorders, Vol 24, Iss 1, Pp 1-9 (2023)
Subject
Hip-knee-ankle angle
Gait analysis
3D knee kinematics
Knee osteoarthritis
Knee surgery
Diseases of the musculoskeletal system
RC925-935
Language
English
ISSN
1471-2474
Abstract
Abstract Background The lower limb mechanical axis was used to assess the severity of knee osteoarthritis (KOA) with varus/valgus deformity and the accuracy of targeted lower limb alignment correction after operation by conventional X-rays. There are lots of parameters to assess the gait in elder patients such as velocity, stride length, step width and swing/stance ratio by knee joint movement analysis system. However, the correlation between the lower limb mechanical axis and gait parameters is not clear. This study is aimed at obtaining the accuracy of the lower limb mechanical axis by the knee joint movement analysis system and the correlation between the lower limb mechanical axis and gait parameters. Methods We analysed 3D knee kinematics during ground gait of 99 patients with KOA and 80 patients 6 months after the operations with the vivo infrared navigation 3D portable knee joint movement analysis system (Opti-Knee®, Innomotion Inc, Shanghai, China). The HKA (Hip-Knee-Ankle) value was calculated and compared to X-ray findings. Results HKA absolute variation after the operation was 0.83 ± 3.76°, which is lower than that before the operation (5.41 ± 6.20°, p = 0.001) and also lower than the entire cohort (3.36 ± 5.72). Throughout the cohort, a significant correlation with low coefficients (r = -0.19, p = 0.01) between HKA value and anterior-posterior displacement was found. In comparing the HKA values measured on the full-length alignment radiographs and 3D knee joint movement analysis system (Opti-Knee), there was a significant correlation with moderate to high coefficients (r = 0.784 to 0.976). The linear correlation analysis showed that there was a significant correlation between the values of HKA measured by X-ray and movement analysis system (R2 = 0.90, p