Real-Time Shear Wave versus Transient Elastography for Predicting Fibrosis: Applicability, and Impact of Inflammation and Steatosis. A Non-Invasive Comparison.
Document Type
article
Source
PLoS ONE, Vol 11, Iss 10, p e0163276 (2016)
Subject
Medicine
Science
Language
English
ISSN
1932-6203
Abstract
Real-time shear wave elastography (2D-SWE) is a two-dimensional transient elastography and a competitor as a biomarker of liver fibrosis in comparison with the standard reference transient elastography by M probe (TE-M). The aims were to compare several criteria of applicability, and to assess inflammation and steatosis impact on elasticity values, two unmet needs.We took FibroTest as the fibrosis reference and ActiTest and SteatoTest as quantitative estimates of inflammation and steatosis. After standardization of estimates, analyses used curve fitting, quantitative Lin concordance coefficient [LCC], and multivariate logistic regression.A total of 2,251 consecutive patients were included. We validated the predetermined 0.2 kPa cut-off as a too low minimal elasticity value identifying not-reliable 2D-SWE results (LCC with FibroTest = 0.0281[-0.119;0.175]. Other criteria, elasticity CV, body mass index and depth of measures were not sufficiently discriminant. The applicability of 2D-SWE (95%CI) 89.6%(88.2-90.8), was significantly higher than that of TE, 85.6%(84.0-87.0; P