Influence of ultrasound transmit frequency on measurement of global longitudinal strain on 2D speckle tracking echocardiography
Document Type
article
Source
Scientific Reports, Vol 13, Iss 1, Pp 1-9 (2023)
Subject
Medicine
Science
Language
English
ISSN
2045-2322
Abstract
Abstract The reproducibility of longitudinal strain measured by 2D speckle tracking echocardiography (2DSTE) may be affected by ultrasound settings. This study investigated the effect of transmit ultrasound frequency on global longitudinal strain (GLS) by 2DSTE. Apical, 2- and 4-chamber, and long-axis views were obtained in consecutive 162 patients using Philips ultrasound devices. Three different frequency presets were used sequentially: high resolution (HRES, 1.9 to 2.1 MHz), general (HGEN, 1.6 to 1.8 MHz), and penetration mode (HPEN, 1.3 to 1.6 MHz). GLS values were determined for each preset using the Philips Q-station software, resulting in GLS-HRES, GLS-HGEN, and GLS-HPEN. Among the 151 patients with successfully measured GLS, a significant difference in GLS was observed among the three presets (p