Improving sensitivity of a BEGe-based high-purity germanium spectrometer through pulse shape analysis
Document Type
article
Source
European Physical Journal C: Particles and Fields, Vol 78, Iss 5, Pp 1-11 (2018)
Subject
Astrophysics
QB460-466
Nuclear and particle physics. Atomic energy. Radioactivity
QC770-798
Language
English
ISSN
1434-6044
1434-6052
Abstract
Abstract We performed Pulse Shape Analysis to separate single-scattered gamma energy deposition events from multiple-scattered photons in a high-sensitivity $$\gamma $$ γ -ray spectrometer. The spectrometer is based on a Broad Energy High Purity Germanium detector and the developed technique uses multivariate analysis by an application of the Multi-Layer Perceptron Neural Network. A very good separation of the single-site- and multi-site events was achieved leading to a significant reduction of the background level of the investigated spectrometer – the double escape peak, rich in single-site events, was reduced by 95%, while the full energy peaks lost at most 25% of their counts. The peak to Compton ratio, calculated for the 2614.5 keV gamma line from $$^{208}$$ 208 Tl, was improved by 114.3%.