CT dose management for neurologic events in patients with cardiac devices: Radiation exposure variation in patients with cardiac devices
Document Type
article
Source
Diagnostic and Interventional Radiology, Vol 28, Iss 1, Pp 98-102 (2022)
Subject
Medical physics. Medical radiology. Nuclear medicine
R895-920
Language
English
ISSN
1305-3825
1305-3612
Abstract
PURPOSE:To compare the inter-center cranial computed tomography (CT) acquisition rates, CT findings, CT related radiation dose, and variability of CT acquisition parameters for neurologic events among patients with implantable cardioverter-defibrillator (ICD) or left ventricular assist device (LVAD).METHODS:A total of 224 patients [ICD group (n = 155) and LVAD group (n = 69)] who had at least one cranial CT scan were enrolled from three medical centers. The variability and effect of the number, indication, and findings of cranial CT scans as well as CT acquisition parameters including tube potential, tube current, tube rotation time (TI), slice collimation (cSL), and spiral or sequential scanning techniques on CT dose index volume (CTDIvol), total dose length product (DLP) were analyzed.RESULTS:The mean DLP value of Center A and mean CTDIvol values of Center A and C were significantly lower than Center B (p < 0.001). The mean CTDIvol and DLP values in the ICD group were substantially lower than the LVAD group (p