Capture threshold of bipolar and unipolar pacing of left ventricle via coronary sinus branch: longitudinal study
Document Type
article
Source
Frontiers in Cardiovascular Medicine, Vol 10 (2023)
Subject
quadripolar lead
pacing capture threshold energy
pacing polarity
steroid eluting lead
resynchronisation therapy
battery longevity
Diseases of the circulatory (Cardiovascular) system
RC666-701
Language
English
ISSN
2297-055X
Abstract
IntroductionThe aim of this paper is to first monitor the changes in the capture threshold of endovascularly placed leads for left ventricle pacing, second to compare the pacing configurations, and third to verify the effect of Steroid elution for endovascular leads.Sample and MethodThe study included 202 consecutive single centre patients implanted with the Quartet™ lead (St. Jude Medical). The capture threshold and related lead parameters were tested during implantation, on the day of the patient’s discharge, and 3, 9, and 15 months after implantation. The electrical energy corresponding to the threshold values for inducing ventricular contraction was recorded for subgroups of patients with bipolar and pseudo-unipolar pacing vectors and electrodes equipped with and without a slow-eluting steroids. The best setting for the resynchronization effect was generally chosen. Capture threshold was taken as a selection criterion only if there were multiple options with (expected) similar resynchronization effect.Results and DiscussionThe measurements showed that the ratio of threshold energies of UNI vs. BI was 5× higher (p