Asynchronous calibration of quantitative computed tomography bone mineral density assessment for opportunistic osteoporosis screening: phantom-based validation and parameter influence evaluation
Document Type
article
Source
Scientific Reports, Vol 12, Iss 1, Pp 1-9 (2022)
Subject
Medicine
Science
Language
English
ISSN
2045-2322
Abstract
Abstract Asynchronous calibration could allow opportunistic screening based on routine CT for early osteoporosis detection. In this phantom study, a bone mineral density (BMD) calibration phantom and multi-energy CT (MECT) phantom were imaged on eight different CT scanners with multiple tube voltages (80–150 kVp) and image reconstruction settings (e.g. soft/hard kernel). Reference values for asynchronous BMD estimation were calculated from the BMD-phantom and validated with six calcium composite inserts of the MECT-phantom with known ground truth. Relative errors/changes in estimated BMD were calculated and investigated for influence of tube voltage, CT scanner and reconstruction setting. Reference values for 282 acquisitions were determined, resulting in an average relative error between calculated BMD and ground truth of − 9.2% ± 14.0% with a strong correlation (R 2 = 0.99; p