Oximetry neither to prescribe long-term oxygen therapy nor to screen for severe hypoxaemia
Document Type
article
Source
ERJ Open Research, Vol 7, Iss 4 (2021)
Subject
Medicine
Language
English
ISSN
2312-0541
23120541
Abstract
Background and objective Transcutaneous pulse oximetry saturation (SpO2) is widely used to diagnose severe hypoxaemia and to prescribe long-term oxygen therapy (LTOT) in COPD. This practice is not based on evidence. The primary objective of this study was to determine the accuracy (false positive and false negative rates) of oximetry for prescribing LTOT or for screening for severe hypoxaemia in patients with COPD. Methods In a cross-sectional study, we correlated arterial oxygen saturation (SaO2) and SpO2 in patients with COPD and moderate hypoxaemia (n=240) and calculated the false positive and false negative rates of SaO2 at the threshold of ≤88% to identify severe hypoxaemia (arterial oxygen tension (PaO2) ≤55 mmHg or PaO2 88% (i.e., false negative result). Conversely, LTOT would be prescribed on the basis of a SaO2 ≤88% in 2% of patients who would not qualify for LTOT (i.e., false positive result). Using a screening threshold of ≤92%, 5% of severely hypoxaemic patients would not be referred for further evaluation. Conclusions Several patients who qualify for LTOT would be denied treatment using a prescription threshold of saturation ≤88% or a screening threshold of ≤92%. Prescription of LTOT should be based on PaO2 measurement.