학술논문

Repeatability and Reproducibility of Retinal Fractal Dimension Measured with Swept-Source Optical Coherence Tomography Angiography in Healthy Eyes: A Proof-of-Concept Study
Document Type
article
Source
Diagnostics, Vol 12, Iss 7, p 1769 (2022)
Subject
fractal dimension
healthy volunteers
swept-source OCT angiography
superficial retinal capillary plexus
deep retinal capillary plexus
retina
Medicine (General)
R5-920
Language
English
ISSN
2075-4418
Abstract
The retinal vascular network fractal dimension (FD) could be a promising imaging biomarker. Our objective was to evaluate its repeatability and reproducibility in healthy eyes. A cross-sectional study was undertaken with young, healthy volunteers who had no reported cardiac risk factors or ocular disease history. For each participant, three SS-OCTA images (12 × 12 mm) were acquired using the Plex Elite 9000 (Carl Zeiss Meditec AG, Jena, Germany) by two ophthalmologists. Automated segmentation was obtained from both the superficial and deep capillary plexuses. FD was estimated by box counting. The intraclass correlation coefficients (ICC) were used as measures for repeatability and reproducibility. A total of 43 eyes of healthy volunteers were included. The mean ± standard deviation (SD) age was 30 ± 6.2 years. The results show good repeatability. The ICC was 0.722 (95% CI, 0.541–0.839) in the superficial capillary plexus and 0.828 (95% CI, 0.705–0.903) in the deep capillary plexus. For reproducibility, the ICC was 0.651 (95% CI, 0.439–0.795) and 0.363 (95% CI, 0.073–0.596) at the superficial and deep capillary plexus, respectively. In this study, the FD of the vascular network measured via SS-OCTA showed good repeatability and reproducibility in healthy participants.