학술논문

Alternative instrument for the evaluation of handgrip strength in Duchenne muscular dystrophy
Document Type
article
Source
BMC Pediatrics, Vol 22, Iss 1, Pp 1-7 (2022)
Subject
Dynamometer
Handgrip strength
Duchenne muscular dystrophy
Pediatrics
RJ1-570
Language
English
ISSN
1471-2431
Abstract
Abstract Background The commonly used dynamometers can be ineffective in evaluating handgrip in patients with Duchenne muscular dystrophy (DMD), especially children with generalized muscle weakness. The aim of this study was to analyze whether the modified sphygmomanometer is an effective instrument for handgrip strength evaluation in patients with DMD, during different stages of the disease. Method The handgrip strength of 33 patients was evaluated by the Jamar dynamometer and the modified sphygmomanometer. Motor function was evaluated by the Motor Function Measurement (MFM) scale. Four evaluations, with a six-month interval between each, were performed: Evaluation 1 (N = 33), Evaluation 2 (N = 24), Evaluation 3 (N = 15), and Evaluation 4 (N = 8). A linear regression model with mixed effects was used for the longitudinal data and descriptive analysis of strength for all four evaluations. Result The first evaluation data presented very high correlations between the dynamometer and the modified sphygmomanometer (r = 0.977; p 0.05). Null values were obtained only when using the dynamometer device. Conclusion The modified sphygmomanometer seems to be more suitable than the dynamometer for measuring handgrip strength in all stages of DMD.