Powered bone marrow biopsy procedures produce larger core specimens, with less pain, in less time than with standard manual devices
Document Type
article
Author
Source
Hematology Reports, Vol 3, Iss 1, Pp e8-e8 (2011)
Subject
Language
English
ISSN
2038-8322
2038-8330
2038-8330
Abstract
Bone marrow sampling remains essential in the evaluation of hematopoietic and many non-hematopoietic disorders. One common limitation to these procedures is the discomfort experienced by patients. To address whether a Powered biopsy system could reduce discomfort while providing equivalent or better results, we performed a randomized trial in adult volunteers. Twenty-six subjects underwent bilateral biopsies with each device. Core samples were obtained in 66.7% of Manual insertions; 100% of Powered insertions (P=0.002). Initial mean biopsy core lengths were 11.1±4.5 mm for the Manual device; 17.0±6.8 mm for the Powered device (P