Cepstral Analysis for Scoring the Quality of Electrocardiograms for Heart Rate Variability
Document Type
article
Source
Frontiers in Physiology, Vol 13 (2022)
Subject
heart rate variability
fourier transform
EKG
power cepstrum
signal quality
premature beat
Physiology
QP1-981
Language
English
ISSN
1664-042X
Abstract
Mobile-health solutions based on heart rate variability often require electrocardiogram (ECG) recordings by inexperienced operators or real-time automatic analyses of long-term recordings by wearable devices in free-moving individuals. In this context, it is useful to associate a quality index with the ECG, scoring the adequacy of the recording for heart rate variability to identify noise or arrhythmias. Therefore, this work aims to propose and validate a computational method for assessing the adequacy of single-lead ECGs for heart rate variability analysis that may run in real time on wearable systems with low computational power. The method quantifies the ECG pseudo-periodic structure employing cepstral analysis. The cepstrum (spectrum of log-spectrum) is estimated on a running ECG window of 10 s before and after “liftering” (filtering in the cepstral domain) to remove slower noise components. The ECG periodicity generates a dominant peak in the liftered cepstrum at the “quefrency” of the mean cardiac interval. The Cepstral Quality Index (CQI) is the ratio between the cepstral-peak power and the total power of the unliftered cepstrum. Noises and arrhythmias reduce the relative power of the cepstral peak decreasing CQI. We analyzed a public dataset of 6072 single-lead ECGs manually classified in normal rhythm or inadequate for heart rate variability analysis because of noise or atrial fibrillation, and the CQI = 47% cut-off identified the inadequate recordings with 79% sensitivity and 85% specificity. We showed that the performance is independent of the lead considering a public dataset of 1,000 12-lead recordings with quality classified as “acceptable” or “unacceptable” by visual inspection. Thus, the cepstrum describes the ECG periodic structure effectively and concisely and CQI appears to be a robust score of the adequacy of ECG recording for heart rate variability analysis, evaluable in real-time on wearable devices.