Effects of Positioning and Notching of Resurfaced Femurs on Femoral Neck Strength: A Biomechanical Test
Document Type
article
Source
Journal of Orthopaedic Surgery, Vol 17 (2009)
Subject
Orthopedic surgery
RD701-811
Language
English
ISSN
2309-4990
23094990
Abstract
Purpose. To assess the effects of positioning and notching of resurfaced femurs on the mechanical strength of third-generation saw bone (TGSB) femurs using an in vitro analogue bone model. Methods. 30 TGSB femurs were equally divided into 6 resurfaced femur groups (intact, anatomic, varus, valgus, anatomically notched, and valgus notched) for testing the load to failure, stiffness, and total energy. Results. Compared to the intact femurs, the load to failure in all resurfaced femurs was significantly decreased by 29 to 57%. Among the resurfaced femurs, valgus and anatomic femurs had the highest load to failure, followed by valgus notched, varus, and anatomically notched femurs. Notching weakened the construct by a further 24 to 30%. Conclusion. To minimise the risk of femoral neck fracture, resurfaced femoral heads should be placed in an anatomic or valgus orientation, and the superior cortex of the femoral neck should remain intact.