Analytical and clinical evaluations of SNIBE Maglumi chemiluminescent immunoassay for the detection of SARS-CoV-2 antigen in salivary samples
Document Type
research-article
Source
Clinical Chemistry and Laboratory Medicine (CCLM). 62(3):572-577
Subject
SARS-CoV-2 antigen
saliva
COVID-19
chemiluminescent assay
Maglumi
Language
English
ISSN
1434-6621
1437-4331
Abstract
Objectives In this study, we describe the analytical and clinical performances of the SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay (MAG-CLIA) on salivary samples. Methods Limit of detection (LOD), linearity and precision were tested for values close to or below the declared LOD. Clinical performance of MAG-CLIA was evaluated on leftover salivary samples from the healthcare workers (HCW) surveillance program, at the University-Hospital of Padova. Salivary samples were analyzed by Lumipulse G SARS-CoV-2 Ag, and in case where the values exceeded 0.41 ng/L, further testing was conducted using TaqPathTM COVID-19 RT-PCR (Applied Biosystems, Thermo Fisher Scientific). Results The estimated MAG-CLIA LOD was 3 ng/L, with repeatability of 7.5 %. Good linearity was demonstrated by diluting two samples at 52.7 ng/L and 211.4 ng/L. Of the 228 HCW samples, 59/228 (25.9 %) were positive, 169/228 (74.1 %) were negative. MAG-CLIA SARS-CoV-2 sAg median level (and interquartile range [IQR]) was 5.03 ng/L ( Conclusions The analytical performances of MAG-CLIA are satisfactory, also when values below LOD were tested. In saliva samples, although specificity was elevated, clinical performance was not comparable with that on nasopharyngeal swabs (NPS).