Measurement of factor XIII (FXIII) activity by an automatic ammonia release assay using iodoacetamide blank-procedure: no more overestimation in the low activity range and better detection of severe FXIII deficiencies
Document Type
research-article
Source
Clinical Chemistry and Laboratory Medicine (CCLM). 54(5):805-809
Subject
ammonia release
blank
deficiencies
factor XIII
iodoacetamide
General Clinical Chemistry and Laboratory Medicine
Language
English
ISSN
1437-4331
1434-6621
Abstract
Background: Laboratory investigation with specific factor XIII (FXIII) assays plays a crucial role in diagnosis of FXIII deficiency. According to the International Society on Thrombosis and Hemostasis (ISTH), it is necessary a blank sample with iodoacetamide, provided by the kit or locally prepared, when the ammonia release assays are used, to avoid FXIII activity overestimation.Methods: In this study we set up a modification of the Berichrom FXIII chromogenic assay, in which iodoacetamide was added by the BCS analyzer in the reaction mixture of the blank sample, without modifications of the original reagents. We analyzed 100 plasma samples of outpatients with clinical symptoms suggestive of a bleeding diathesis (20 samples had FXIII activity Results: In all samples blank subtraction significantly reduced FXIII activity, mostly in the low activity range group (from 10.1% to 2.4%, pConclusions: Despite the low number of samples included in the study, the described automatic procedure seemed to decrease FXIII activity overestimation and, especially for low activity range samples (