Analytical evaluation of the Nittobo Medical tartrate resistant acid phosphatase isoform 5b (TRACP-5b) EIA and comparison with IDS iSYS in different clinically defined populations
Document Type
research-article
Source
Clinical Chemistry and Laboratory Medicine (CCLM). 60(3):394-400
Subject
analytical validation
bone resorption
bone turnover marker
harmonization
tartrate resistant acid phosphatase isoform 5b
TRACP-5b
Language
English
ISSN
1434-6621
1437-4331
Abstract
Objectives Tartrate-resistant acid phosphatase, isoform 5b (TRACP-5b) is a bone resorption marker not influenced by renal function or food intake. TRACP-5b can be measured with Nittobo Medical enzymatic-immunoassay and IDS-iSYS automated immunoassay. We evaluated the Nittobo assay and established reference ranges for a Western-European population. We compared Nittobo and IDS results in different well-defined clinical populations. Methods We established the limits of detection and quantification (LOD-LOQ), linearity, imprecision and the reference ranges in 119 males, 50 women (60 years) for TRACP-5b with the Nittobo assay. We compared both assays in 30 hemodialyzed (HD), and 40 stage 3–5 patients suffering from chronic kidney disease (CKD), 40 patients suffering from rheumatoid arthritis and osteoporosis and 80 post-menopausal women. We measured TRACP-5b, β-crosslaps (β-CTX), bone alkaline phosphatase (B-ALP) and PTH in 20 hemodialyzed (HD) and 40 CKD patients. Results LOD and LOQ were 0.02 and 0.35 U/L. CV ranged from 8.3 to 4.3% (2/5 samples presenting CV > desirable CV). Method was linear up to of 11.3 U/L. Upper and lower limits of normality were 0.8–7.6 U/L in men, 0.9–4.7 U/L in women 60. The regression equation between the 2 methods was Nittobo = 1.13 (95% CI: 1.09–1.16) × iSYS − 0.4 (95% CI: −0.5; −0.3). TRACP-5b and b-ALP were in their respective reference ranges for most of CKD and HD patients. That was not the case for β-CTX, which increased with decreasing eGFR. Conclusions Nittobo TRACP-5b presents interesting analytical features and a good concordance with IDS iSYS. These methods could thus potentially be harmonized.