Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading
Document Type
research-article
Source
Clinical Chemistry and Laboratory Medicine (CCLM). 60(8):1202-1210
Subject
laboratory automation
pneumatic tube system
pre-analytical phase
sample transport
Language
English
ISSN
1434-6621
1437-4331
Abstract
Objectives A carrier prototype by Aerocom® (Schwäbisch Gmünd, Germany) for pneumatic tube systems (PTS) is able to transport 9 blood tubes which are automatically fixed by closing the lid. In this study, we examined the influence of the transport on blood sample quality using the carrier prototype comparing to courier transport and a conventional carrier (AD160, Aerocom®). Methods Triplicate blood samples sets (1 lithium heparin, 1 EDTA, 1 sodium citrate) of 35 probands were split among the transportation methods: 1. courier, 2. conventional carrier, and 3. carrier prototype. After transport 51 measurands from clinical chemistry, hematology and coagulation were measured and compared. Results Overall, 49 of the investigated 51 measurands showed a good concordance among the three transport types, especially between the conventional carrier and the carrier prototype. Focusing on well-known hemolysis sensitive measurands, potassium showed no statistically significant differences. However, between courier and both carrier types lactate dehydrogenase (LDH) and free hemoglobin (fHb) showed statistically significant shifts, whereas the clinical impact of the identified differences was neglectable. The median concentration of fHb, for example, was 0.29 g/L (18 µmol/L), 0.31 g/L (19 µmol/L) and 0.32 g/L (20 µmol/L) for courier transport, conventional carrier and carrier prototype, respectively. These differences cannot be resolved analytically since the minimal difference (MD) for fHb is 0.052 g/L (3.23 µmol/L), at this concentration. Conclusions The carrier prototype by Aerocom® is suitable for transportation of diagnostic blood samples. The overall workflow is improved by decreasing hands-on-time on the ward and laboratory while minimizing the risk of incorrectly packed carriers.