Development of a $^{83\mathrm{m}}$Kr source for the calibration of the CENNS-10 Liquid Argon Detector
Document Type
Working Paper
Author
COHERENT Collaboration; Akimov, D.; An, P.; Awe, C.; Barbeau, P. S.; Becker, B.; Belov, V.; Bernardi, I.; Blackston, M. A.; Blokland, L.; Bolozdynya, A.; Cabrera-Palmer, B.; Chen, N.; Chernyak, D.; Conley, E.; Daughhetee, J.; Coello, M. del Valle; Detwiler, J. A.; Durand, M. R.; Efremenko, Y.; Elliott, S. R.; Fabris, L.; Febbraro, M.; Fox, W.; Galindo-Uribarri, A.; Rosso, A. Gallo; Green, M. P.; Hansen, K. S.; Heath, M. R.; Hedges, S.; Hughes, M.; Johnson, T.; Khromov, A.; Konovalov, A.; Kozlova, E.; Kumpan, A.; Li, L.; Librande, J. T.; Link, J. M.; Liu, J.; Mann, K.; Markoff, D. M.; McGoldrick, O.; Mueller, P. E.; Newby, J.; Parno, D. S.; Pentilla, S.; Pershey, D.; Radford, D.; Rapp, R.; Ray, H.; Raybern, J.; Razuvaeva, O.; Reyna, D.; Rich, G. C.; Rudik, D.; Runge, J.; Salvat, D. J.; Scholberg, K.; Shakirov, A.; Simakov, G.; Snow, W. M.; Sosnovtsev, V.; Suh, B.; Tayloe, R.; Tellez-Giron-Flores, K.; Thornton, R. T.; Tolstukhin, I.; Vanderwerp, J.; Varner, R. L.; Venkataraman, R.; Virtue, C. J.; Visser, G.; Wiseman, C.; Wongjirad, T.; Yang, J.; Yen, Y. -R.; Yoo, J.; Yu, C. -H.; Zettlemoyer, J.
Source
JINST 16 P04002 (2021)
Subject
Language
Abstract
We report on the preparation of and calibration measurements with a $^{83\mathrm{m}}$Kr source for the CENNS-10 liquid argon detector. $^{83\mathrm{m}}$Kr atoms generated in the decay of a $^{83}$Rb source were introduced into the detector via injection into the Ar circulation loop. Scintillation light arising from the 9.4 keV and 32.1 keV conversion electrons in the decay of $^{83\mathrm{m}}$Kr in the detector volume were then observed. This calibration source allows the characterization of the low-energy response of the CENNS-10 detector and is applicable to other low-energy-threshold detectors. The energy resolution of the detector was measured to be 9$\%$ at the total $^{83\mathrm{m}}$Kr decay energy of 41.5 keV. We performed an analysis to separately calibrate the detector using the two conversion electrons at 9.4 keV and 32.1 keV
Comment: v2: As accepted to JINST
Comment: v2: As accepted to JINST