학술논문

Dark Count of 20-inch PMTs Generated by Natural Radioactivity
Document Type
Working Paper
Source
Subject
Physics - Instrumentation and Detectors
High Energy Physics - Experiment
Language
Abstract
The primary objective of the JUNO experiment is to determine the ordering of neutrino masses using a 20-kton liquid-scintillator detector. The 20-inch photomultiplier tube (PMT) plays a crucial role in achieving excellent energy resolution of at least 3% at 1 MeV. Understanding the characteristics and features of the PMT is vital for comprehending the detector's performance, particularly regarding the occurrence of large pulses in PMT dark counts. This research paper aims to further investigate the origin of these large pulses in the 20-inch PMT dark count rate through measurements and simulations. The findings confirm that the main sources of the large pulses are natural radioactivity and muons striking the PMT glass. By analyzing the PMT dark count rate spectrum, it becomes possible to roughly estimate the radioactivity levels in the surrounding environment.
Comment: 16 pages, 8 figures, and 5 tables