Calibration of the charge and energy loss per unit length of the MicroBooNE liquid argon time projection chamber using muons and protons
Document Type
Working Paper
Author
MicroBooNE collaboration; Adams, C.; Alrashed, M.; An, R.; Anthony, J.; Asaadi, J.; Ashkenazi, A.; Balasubramanian, S.; Baller, B.; Barnes, C.; Barr, G.; Basque, V.; Bass, M.; Bay, F.; Berkman, S.; Bhanderi, A.; Bhat, A.; Bishai, M.; Blake, A.; Bolton, T.; Camilleri, L.; Caratelli, D.; Terrazas, I. Caro; Carr, R.; Fernandez, R. Castillo; Cavanna, F.; Cerati, G.; Chen, Y.; Church, E.; Cianci, D.; Cohen, E. O.; Conrad, J. M.; Convery, M.; Cooper-Troendle, L.; Crespo-Anadon, J. I.; Del Tutto, M.; Devitt, D.; Diaz, A.; Domine, L.; Duffy, K.; Dytman, S.; Eberly, B.; Ereditato, A.; Sanchez, L. Escudero; Esquivel, J.; Evans, J. J.; Fitzpatrick, R. S.; Fleming, B. T.; Foppiani, N.; Franco, D.; Furmanski, A. P.; Garcia-Gamez, D.; Gardiner, S.; Genty, V.; Goeldi, D.; Gollapinni, S.; Goodwin, O.; Gramellini, E.; Green, P.; Greenlee, H.; Grosso, R.; Gu, L.; Gu, W.; Guenette, R.; Guzowski, P.; Hamilton, P.; Hen, O.; Hill, C.; Horton-Smith, G. A.; Hourlier, A.; Huang, E. C.; Itay, R.; James, C.; de Vries, J. Jan; Ji, X.; Jiang, L.; Jo, J. H.; Johnson, R. A.; Joshi, J.; Jwa, Y. J.; Karagiorgi, G.; Ketchum, W.; Kirby, B.; Kirby, M.; Kobilarcik, T.; Kreslo, I.; Lepetic, I.; Li, Y.; Lister, A.; Littlejohn, B. R.; Lockwitz, S.; Lorca, D.; Louis, W. C.; Luethi, M.; Lundberg, B.; Luo, X.; Marchionni, A.; Marcocci, S.; Mariani, C.; Marshall, J.; Martin-Albo, J.; Caicedo, D. A. Martinez; Mason, K.; Mastbaum, A.; McConkey, N.; Meddage, V.; Mettler, T.; Miller, K.; Mills, J.; Mistry, K.; Mohayai, T.; Mogan, A.; Moon, J.; Mooney, M.; Moore, C. D.; Mousseau, J.; Murphy, M.; Murrells, R.; Naples, D.; Neely, R. K.; Nienaber, P.; Nowak, J.; Palamara, O.; Pandey, V.; Paolone, V.; Papadopoulou, A.; Papavassiliou, V.; Pate, S. F.; Paudel, A.; Pavlovic, Z.; Piasetzky, E.; Porzio, D.; Prince, S.; Pulliam, G.; Qian, X.; Raaf, J. L.; Rafique, A.; Ren, L.; Rochester, L.; Rogers, H. E.; Ross-Lonergan, M.; von Rohr, C. Rudolf; Russell, B.; Scanavini, G.; Schmitz, D. W.; Schukraft, A.; Seligman, W.; Shaevitz, M. H.; Sharankova, R.; Sinclair, J.; Smith, A.; Snider, E. L.; Soderberg, M.; Soldner-Rembold, S.; Soleti, S. R.; Spentzouris, P.; Spitz, J.; Stancari, M.; John, J. St.; Strauss, T.; Sutton, K.; Sword-Fehlberg, S.; Szelc, A. M.; Tagg, N.; Tang, W.; Terao, K.; Thornton, R. T.; Toups, M.; Tsai, Y. -T.; Tufanli, S.; Usher, T.; Van De Pontseele, W.; Van de Water, R. G.; Viren, B.; Weber, M.; Wei, H.; Wickremasinghe, D. A.; Williams, Z.; Wolbers, S.; Wongjirad, T.; Woodruff, K.; Wospakrik, M.; Wu, W.; Yang, T.; Yarbrough, G.; Yates, L. E.; Zeller, G. P.; Zennamo, J.; Zhang, C.
Source
JINST 15, P03022 (2020)
Subject
Language
Abstract
We describe a method used to calibrate the position- and time-dependent response of the MicroBooNE liquid argon time projection chamber anode wires to ionization particle energy loss. The method makes use of crossing cosmic-ray muons to partially correct anode wire signals for multiple effects as a function of time and position, including cross-connected TPC wires, space charge effects, electron attachment to impurities, diffusion, and recombination. The overall energy scale is then determined using fully-contained beam-induced muons originating and stopping in the active region of the detector. Using this method, we obtain an absolute energy scale uncertainty of 2\% in data. We use stopping protons to further refine the relation between the measured charge and the energy loss for highly-ionizing particles. This data-driven detector calibration improves both the measurement of total deposited energy and particle identification based on energy loss per unit length as a function of residual range. As an example, the proton selection efficiency is increased by 2\% after detector calibration.
Comment: Accepted version
Comment: Accepted version