Precision-Optimization of Wavelengths in Diode-Array Detection in Separation Science
Document Type
Article
Source
Analytical Sciences; December 1997, Vol. 13 Issue: 6 p951-962, 12p
Subject
Language
ISSN
09106340; 13482246
Abstract
With the theoretically predicted relative standard deviation (RSD) of measurements as a criterion, the ultra-violet detection wavelengths are optimized in a capillary electrophoresis system equipped with a diode-array, when analyte signals are baseline-separated in the time-direction and the measurements are the integration results in the time and wavelength directions. Some cold drugs and additives are taken as examples. The auto-correlation of an instrumental baseline at a diode (time-correlation) and the cross-correlation of measurements between different diodes (wavelength-correlation) are both taken into account for a better prediction of RSD. A simple method for estimating the wavelength-correlation is presented with a theoretical proof. This paper also shows that the measurement precision is governed to a large extent by the zero level setting of the time integration which accumulates the instrumental output relative to the zero level. The optimum (minimum RSD) is selected from among the wavelengths ( 190 - 300 nm) and possible zero level settings. The validity of the RSD prediction is statistically verified.