Direct Digital Radiography Versus Storage Phosphor Radiography in the Detection of Wrist Fractures
Document Type
Article
Source
Clinical Radiology; April 2002, Vol. 57 Issue: 4 p258-262, 5p
Subject
Language
ISSN
00099260; 1365229X
Abstract
AIM: To define the value of digital radiography with a clinical flat panel detector system for evaluation of wrist fractures in comparison with state of the art storage phosphor radiography. MATERIAL AND METHODS: Hard copy images of 26 fractured wrist specimens were acquired with the same exposure dose on a state of the art storage phosphor radiography system and a clinical flat panel detector. Image features like cortical bone surface, trabecular bone, soft tissues and fracture delineation were independently analysed by 4 observers using a standardised protocol. Image quality ratings were evaluated with an analysis of variance (ANOVA). RESULTS: Flat panel detector radiographs were rated superior with respect to cortical and trabecular bone representation as well as fracture evaluation, while storage phosphor radiographs produced better soft tissue detail. CONCLUSION: In some of the observed image quality aspects, the performance of caesium iodide/amorphous silicon flat panel detector exceeds state of the art storage phosphor radiography. This makes it well suited for skeletal imaging particularly in trauma as seen in the detection of wrist fractures. Peer, S. et al. (2002). Clinical Radiology57, 258–262.