Vacuum ultraviolet radiometry. 3: The argon mini-arc as a new secondary standard of spectral radiance
Document Type
Article
Source
Applied Optics; February 1977, Vol. 16 Issue: 2 p367-376, 10p
Subject
Language
ISSN
1559128X; 21553165
Abstract
A miniature argon arc has been designed and tested as a new transfer standard of spectral radiance for the wavelength range from 114 nm to 330 nm. Calibration has been performed using two primary standard sources: the hydrogen arc from 130 nm to 330 nm and the blackbody line radiator from 114 nm to 130 nm. The mini-arc provides an intense, stable, and reproducible uv continuum with dc power requirements of less than 1.5 kW. The arc characteristics have been investigated, and the sensitivity of the radiant power output to various operating parameters has been measured. The rms uncertainty in the spectral radiance is estimated to be 5.3% above 140 nm and 10.1% between 114 nm and 140 nm, due primarily to uncertainties in the primary standard sources.