학술논문

Decreased radiation exposure using pulsed fluoroscopy and a detachable pedicle marker and probe to place pedicle screws: a comparison to current fluoroscopy techniques and CT navigation
Document Type
Article
Source
Spine Deformity; 20240101, Issue: Preprints p1-7, 7p
Subject
Language
ISSN
2212134X; 22121358
Abstract
Study design: Quality improvement evaluation with retrospective analysis. Objectives: To compare a technique to place pedicle screws (PS) using a novel detachable pedicle marker and probe (DPMP) and pulsed fluoroscopy (PF) vs. conventional technique utilizing PF with standard instruments (SI) and O-arm. Summary of background data: Spinal fusion with pedicle screw instrumentation (PSI) is the mainstay in treatment of spinal deformities. Reports suggest that CT navigated (O-arm) PS placement is more accurate than fluoroscopy. However, these studies have not considered the increased radiation exposure associated with CT. Methods: Thirty-six patients with spinal deformity had PSI using PF and DPMPs. Accuracy of PS placement and radiation data from 14 dosimeters placed on the patient and around the operating room was analyzed. Results were compared to published data. Results: Mean fluoroscopic time was 13.4 s (range 6.0–32.4), and the mean cumulative dose was 3.1 mGy (range 0.2–16.4). Median estimated effective dose to the patient was 0.22 mSv (range 0.0–0.7). The effective dose of radiation was reduced by 80% (0.22 mSv vs. 1.11 mSv) compared to low-dose O-arm. The surgical team did not receive any detectable radiation. The seconds of PF used to assist and confirm placement of PSs was reduced to 1.2 s/level compared to previous reports of 4.49 s/level using SIs. DPMPs reduced fluoroscopy to 0.84 s/PS compared to 7.36 s/PS using SIs to assist and confirm PS placement. PSs were accurately placed in 561 of 576 (97.4%), which is comparable to O-arm and fluoroscopy with SIs. Conclusions: PS placement using PF and DPMPs to assist and confirm PS placement lowers radiation exposure to the patient and surgical team without compromising accuracy compared to O-arm and fluoroscopy with SIs. Level of evidence: Therapeutic, Level IV (Retrospective case series, historical control).