Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer.
Document Type
Article
Source
American Journal of Clinical Pathology; January 2000, Vol. 113 Issue: 1 p75-86, 12p
Subject
Language
ISSN
00029173; 19437722
Abstract
Thirty drugs used primarily in critical care and hospital settings were tested in vitro to observe interference on glucose measurements with 6 hand-held glucose meters and a portable glucose analyzer. Paired differences of glucose measurements between drug-spiked samples and unspiked control samples were calculated to determine bias. A criterion of +/- 6 mg/dL was used as the cutoff for interference. Ascorbic acid interfered with the measurements on all glucose devices evaluated. Acetaminophen, dopamine, and mannitol interfered with glucose measurements on some devices. Dose-response relationships help assessment of drug interference in clinical use. High dosages of these drugs may be given to critically ill patients or self-administered by patients without medical supervision. Package inserts for the glucose devices may not provide adequate warning information. Hence, we recommend that clinicians choose glucose devices carefully and interpret results cautiously when glucose measurements are performed during or after drug interventions.