Chemiluminescence underestimates nitric oxide concentration in the presence of potent inhalation anaesthetics.
Document Type
Article
Source
BJA: British Journal of Anaesthesia; May 2000, Vol. 84 Issue: 5 p596-599, 4p
Subject
Language
ISSN
00070912; 14716771
Abstract
We investigated the effect of potent inhalation anaesthetics on nitric oxide (NO) concentration measured by the chemiluminescence method. We found that the NO concentration was increasingly underestimated with increasing concentrations of halothane, isoflurane, enflurane and sevoflurane (r2 = 0.918-0.997, P < 0.01). Statistical analysis showed that the four inhalation agents at the same concentration produced a similar error in the measured NO concentration. In the presence of a fixed concentration of sevoflurane (5.0%), isoflurane (5.2%), enflurane (4.5%) or halothane (6.1%), the rate of reduction in the measured NO concentration increased in proportion to the NO concentration (r2 = 0.909-0.982, P < 0.01). No direct chemical interaction between the potent inhalation agents and NO was detected by gas chromatography-mass spectrometry. We conclude that NO concentration can be underestimated when measured by the chemiluminescence method in the presence of potent inhalation agents. This underestimation may result from emission absorption and/or the quenching phenomenon, but is not attributable to a chemical reaction between the inhalation agent and NO.