A line imaging system for measurement of eyelid movements.
Document Type
Academic Journal
Author
Gittins J; Medical Physics Department, Leicester Royal Infirmary, UK.; Martin KSheldrick JH
Source
Publisher: IOP Pub. Ltd Country of Publication: England NLM ID: 9306921 Publication Model: Print Cited Medium: Print ISSN: 0967-3334 (Print) Linking ISSN: 09673334 NLM ISO Abbreviation: Physiol Meas Subsets: MEDLINE
Subject
Language
English
ISSN
0967-3334
Abstract
A line scan imaging system which provides a facility for comprehensive measurement and analysis of eyelid motion in human subjects is described. The device obtains parasagittal line images of the eye and eyelids by focusing an image of the eye onto a linear 256-element CCD array. Line images are acquired and digitized at a rate of 200 per second and stored in a buffer memory. The stored data are subsequently analysed on a PC to give measurements of eyelid displacement and velocity. Measurements of eyelid displacement and velocity during normal blinks in five subjects were comparable with published results from other measurement techniques.