Reliability and performance of the IRRAflow® system for intracranial lavage and evacuation of hematomas-A technical note.
Document Type
Academic Journal
Author
Haldrup M; Department of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark.; Nazari M; Department of Engineering, Electrical and Computer Engineering, Aarhus University, Aarhus N, Denmark.; Gu C; Department of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark.; Rasmussen M; Department of Clinical Medicine, Aarhus University, Aarhus N, Denmark.; Department of Anesthesiology, Section of Neuroanesthesia, Aarhus University Hospital, Aarhus, Denmark.; Dyrskog S; Department of Intensive Care, Aarhus University Hospital, Aarhus N, Denmark.; Ziegler Simonsen C; Department of Intensive Care, Aarhus University Hospital, Aarhus N, Denmark.; Department of Neurology, Aarhus University Hospital, Aarhus N, Denmark.; Grønhøj M; Department of Neurosurgery, Odense University Hospital, Odense, Denmark.; Poulsen FR; Department of Neurosurgery, Odense University Hospital, Odense, Denmark.; Ur Rehman N; Department of Engineering, Electrical and Computer Engineering, Aarhus University, Aarhus N, Denmark.; Rosendal Korshoej A; Department of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark.; Department of Clinical Medicine, Aarhus University, Aarhus N, Denmark.
Source
Publisher: Public Library of Science Country of Publication: United States NLM ID: 101285081 Publication Model: eCollection Cited Medium: Internet ISSN: 1932-6203 (Electronic) Linking ISSN: 19326203 NLM ISO Abbreviation: PLoS One Subsets: MEDLINE
Subject
Language
English
Abstract
Background: Intraventricular hemorrhage (IVH) is a severe condition with poor outcomes and high mortality. IRRAflow® (IRRAS AB) is a new technology introduced to accelerate IVH clearance by minimally invasive wash-out. The IRRAflow® system performs active and controlled intracranial irrigation and aspiration with physiological saline, while simultaneously monitoring and maintaining a stable intracranial pressure (ICP). We addressed important aspects of the device implementation and intracranial lavage.
Method: To allow versatile investigation of multiple device parameters, we designed an ex vivo lab setup. We evaluated 1) compatibility between the IRRAflow® catheter and the Silverline f10 bolt (Spiegelberg), 2) the physiological and hydrodynamic effects of varying the IRRAflow® settings, 3) the accuracy of the IRRAflow® injection volumes, and 4) the reliability of the internal ICP monitor of the IRRAflow®.
Results: The IRRAflow® catheter was not compatible with Silverline bolt fixation, which was associated with leakage and obstruction. Design space exploration of IRRAflow® settings revealed that appropriate settings included irrigation rate 20 ml/h with a drainage bag height at 0 cm, irrigation rate 90 ml/h with a drainage bag height at 19 cm and irrigation rate 180 ml/h with a drainage bag height at 29 cm. We found the injection volume performed by the IRRAflow® to be stable and reliable, while the internal ICP monitor was compromised in several ways. We observed a significant mean drift difference of 3.16 mmHg (variance 0.4, p = 0.05) over a 24-hour test period with a mean 24-hour drift of 3.66 mmHg (variance 0.28) in the pressures measured by the IRRAflow® compared to 0.5 mmHg (variance 1.12) in the Raumedic measured pressures.
Conclusion: Bolting of the IRRAflow® catheter using the Medtronic Silverline® bolt is not recommendable. Increased irrigation rates are recommendable followed by a decrease in drainage bag level. ICP measurement using the IRRAflow® device was unreliable and should be accompanied by a control ICP monitor device in clinical settings.