Development and validation of an LC-MS/MS method to quantify lysergic acid diethylamide (LSD), iso-LSD, 2-oxo-3-hydroxy-LSD, and nor-LSD and identify novel metabolites in plasma samples in a controlled clinical trial.
Document Type
Academic Journal
Author
Dolder PC; Laboratory Medicine, University Hospital and University of Basel, Basel, Switzerland.; Division of Clinical Pharmacology and Toxicology, Department of Biomedicine and Department of Clinical Research, University Hospital and University of Basel, Basel, Switzerland.; Liechti ME; Laboratory Medicine, University Hospital and University of Basel, Basel, Switzerland.; Rentsch KM; Laboratory Medicine, University Hospital and University of Basel, Basel, Switzerland.
Source
Publisher: Wiley Country of Publication: United States NLM ID: 8801384 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1098-2825 (Electronic) Linking ISSN: 08878013 NLM ISO Abbreviation: J Clin Lab Anal Subsets: MEDLINE
Subject
Language
English
Abstract
Background: Lysergic acid diethylamide (LSD) is a widely used recreational drug. The aim of this study was to develop and validate a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the quantification of LSD, iso-LSD, 2-oxo-3-hydroxy LSD (O-H-LSD), and nor-LSD in plasma samples from 24 healthy subjects after controlled administration of 100 μg LSD in a clinical trial. In addition, metabolites that have been recently described in in vitro studies, including lysergic acid monoethylamide (LAE), lysergic acid ethyl-2-hydroxyethylamide (LEO), 2-oxo-LSD, trioxylated-LSD, and 13/14-hydroxy-LSD, should be identified.
Methods: Separation of LSD and its metabolites was achieved on a reversed phase chromatography column after turbulent-flow online extraction. For the identification and quantification, a triple-stage quadrupole LC-MS/MS instrument was used.
Results: The validation data showed slight matrix effects for LSD, iso-LSD, O-H-LSD, or nor-LSD. Mean intraday and interday accuracy and precision were 105%/4.81% and 105%/4.35% for LSD, 98.7%/5.75% and 99.4%/7.21% for iso-LSD, 106%/4.54% and 99.4%/7.21% for O-H-LSD, and 107%/5.82% and 102%/5.88% for nor-LSD, respectively. The limit of quantification was 0.05 ng/mL for LSD, iso-LSD, and nor-LSD and 0.1 ng/mL for O-H-LSD. The limit of detection was 0.01 ng/mL for all compounds.
Conclusion: The method described herein was accurate, precise, and the calibration range within the range of expected plasma concentrations. LSD was quantified in the plasma samples of the 24 subjects of the clinical trial, whereas iso-LSD, O-H-LSD, nor-LSD, LAE, LEO, 13/14-hydroxy-LSD, and 2-oxo-LSD could only sporadically be detected but were too low for quantification.