Improved dose localization with dual energy photon irradiation in treatment of lateralized intracranial malignancies.
Document Type
Academic Journal
Author
Cooley G; Medical College of Wisconsin, Department of Radiation Oncology, Milwaukee 53226.; Gillin MTMurray KJWilson JFJanjan NA
Source
Publisher: Elsevier, Inc Country of Publication: United States NLM ID: 7603616 Publication Model: Print Cited Medium: Print ISSN: 0360-3016 (Print) Linking ISSN: 03603016 NLM ISO Abbreviation: Int J Radiat Oncol Biol Phys Subsets: MEDLINE
Subject
Language
English
ISSN
0360-3016
Abstract
Dual energy photon irradiation (6 MV and 20 MV) was compared to conventional treatment planning with 6 MV photons in a lateralized intracranial malignancy. Dose volume analysis was performed of both the tumor plus a 2 cm margin (target volume, TV) and normal tissues (NT). Parallel opposed treatment using weightings of 1:1, 1.5:1, and 2:1 were compared for 6 MV photons alone or in combination with 20 MV photons. Uniform treatment of the TV was accomplished within the 60 Gy isodose. Significant differences were observed, however, in NT volumes receiving greater than or equal to 60 Gy and 45-59 Gy. Dual photon energy reduced treatment of NT volumes to greater than or equal to 60 Gy by 13% (177 cm3 vs 204 cm3 in 2:1 weighting) to 70% (147 cm3 vs 498 cm3 in 1:1 weighting) for comparable plans. Dose optimization was also performed for both 6 MV alone or in combination with 20 MV photons. Usual approaches to achieve dose lateralization with conventional isocentric techniques were applied including parallel opposed 6 MV photons ipsilaterally weighted 3.4:1 (POP), and a 110 degrees arc rotational field used to limit treatment to the eye (ARC). Dual energy photon optimized plans included a three beam parallel opposed plan (TOP) and a mixed photon ipsilateral (IPSI) approach. The technique using parallel opposed 20 MV photons and ipsilateral 6 MV photons (TOP) used beam weightings of 1.1 (contralateral 20 MVX): 1.6 (ipsilateral 6 MVX): 1 (ipsilateral 20 MVX) to achieve dose optimization. The ipsilateral approach with 6 MVX and 20 MVX (IPSI) used beam weightings of 1:1.4, respectively. All optimized plans demonstrated a 41% (120 cm3; POP) to 53% (95 cm3; TOP) improvement over parallel opposed 6 MV photons weighted 2:1 (204 cm3) in NT volume receiving greater than or equal to 60 Gy. Comparison of optimized treatment showed the IPSI plan to be superior, treating 12% of NT volume to greater than or equal to 60 Gy and 38% to 45-59 Gy; the 6 MV POP plan resulted in NT volumes of 15% and 51%, respectively, for those dose levels. Dual photon energy irradiation of lateralized intracranial malignancies allows reduction of dose to normal tissue volumes while achieving excellent coverage of the target volume. Treatment planning should be performed in all lateralized intracranial lesions to achieve dose optimization exploiting depth dose characteristics.