Computational analysis of the effects of interprosthetic distance on normal and reduced cortical thickness femur models.
Document Type
Academic Journal
Author
Güngörürler M; Department of Biomechanics, School of Medicine Health Science Institute, Dokuz Eylul University, Izmir, Turkey.; Gürsan O; Department of Orthopaedics and Traumatology, Dokuz Eylul University, Izmir, Turkey.; Havıtçıoğlu H; Department of Orthopaedics and Traumatology, Dokuz Eylul University, Izmir, Turkey.
Source
Publisher: Sage Country of Publication: England NLM ID: 8908934 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 2041-3033 (Electronic) Linking ISSN: 09544119 NLM ISO Abbreviation: Proc Inst Mech Eng H Subsets: MEDLINE
Subject
Language
English
Abstract
Distal femoral fractures associated with the femoral stem in a well-fixed hip arthroplasty pose a risk of an interprosthetic fracture, the treatment of which is known as difficult. To effectively prevent and treat IP fractures, biomechanical effects must be demonstrated. We defined eight variations of the interprosthetic distance ranging from 48 mm overlap to 128 mm gap. Femoral geometries with normal and reduced cortical thickness were modeled to evaluate the effects of cortical thickness. In addition to the intact model, a total of 16 finite element models were analyzed under physiological boundary conditions. Maximum and minimum principal strains on the lateral and medial cortex surfaces were always found to be greater in models with reduced cortical thickness than in normal femurs. The model with 48 mm overlapping interprosthetic distance produced the least peak strain and the model with 16 mm interprosthetic gap produced the greatest strain with both normal and reduced cortical thickness. The screw holes produced local strain concentrations and increased the peak strains on the cortex surfaces, especially close to the stem tip. Statistically, a significant correlation ( R 2  = 0.9483) was found between strain shielding and interprosthetic distance. Axial stiffness, interfragmentary shear motion, and maximum von-Mises stress on the distal plate showed a high correlation with the interprosthetic distance. It was concluded that the overlapping structures are superior to other fixations we analyzed in that they offer better mechanical stability and eliminates the local strain concentrations.